Page 146 of 188
User Manual MAS-100 Regulus en.docx, Version: 4.0
11) After successful login, the PC Software starts in the Adjustment/Calibration menu.
12) Select the head and the perforated lid on it:
300 × 0.6 or 400 × 0.7
13) Select Manual 3-point.
14) Set the desired Adjustment flow.
Start the process with the button.
Pressing afterwards aborts the process.
Note
The standard factory setting for the Adjustment flow is 100 SLPM.
Note
The standard factory setting for the Adjustment Span is 10 SLPM.
15) Press Yes to proceed.
Do not add or remove instruments.
Pressing No aborts the process.