User Manual MAS-100 Regulus en.docx, Version: 4.0
Adjustment point
16) Change the blower Speed until the flow on the ane-
mometer is stable and close to the Adjust-
ment flow.
The current Adjustment flow is displayed in the
Mean flow box.
Pressing aborts the process.
Note
If a reference value must be recorded for documentation purposes, go to the
Flow Adjustment Data tab in the PC Software and use the ADC at target flow
value after the adjustment.
Annex A.2.1 provides additional background information.
17) Anemometer reading:
Allow the anemometer 40seconds to settle prior taking readings.
Take 3 consecutive readings from the anemometer.
Note these 3 flow values: F
1
, F
2
, F
3
Calculate the Mean flow value: F
m
=(F
1
+F
2
+F
3
)/3
18) Iteration:
- Modify blower Speed
- Take anemometer reading
- Is the F
m
value firmly inside the range of Adjustment flow ±0.5SLPM?
o If No: Redo iteration.
o
If Yes: Press the button and continue.
Example
Adj. Point
Adj. flow 95.0SLPM 105.0SLPM 100.0SLPM
Tolerance ±0.5SLPM
Adj. flow range 94.5 - 95.5SLPM 104.5 - 105.5SLPM 99.5 - 100.5SLPM
19) Enter the last calculated Mean flow value F
m
by
modifying the displayed value.
Press the button to save the value and to pro-
ceed.
Pressing aborts the process.