Page 148 of 188
User Manual MAS-100 Regulus en.docx, Version: 4.0
Adjustment point ❷
20) Proceed as with steps 16-19 with the adjustment flow rate corresponding to Adjust-
ment Point ❷.
Adjustment point ❸
21) Proceed as with steps 16-19 with the adjustment flow rate corresponding to Adjust-
ment Point ❸.
22) Press the Yes button to accept the adjustment and
store the values in the air sampler.
Pressing No aborts the process.
23) If used, remove the MAS-100 Family Calibration Disk or the Petri dish.
24) This procedure may be repeated for all other sampling heads.
25) End of process.