12) After successful login, the PC Software starts in the Adjustment/Calibration menu.
13) Select the head and the perforated lid on it:
300 × 0.6 or 400 × 0.7
14) Select Automatic adjustment.
15) Set the adjustment parameters:
– Adjustment flow
– Adjustment span
Start the process with the button.
Pressing afterwards aborts the process.
Note
The standard factory setting for the Adjustment flow is 100 SLPM.
Note
The standard factory setting for the Adjustment Span is 10 SLPM.
16) Press Yes to proceed.
Do not add or remove instruments.
Pressing No aborts the process.
17) The air sampler will now adjust through all three adjustment points. No action from
the user is required, the process will continue automatically.