18) Press the Yes button to accept the adjustment and
store the values in the air sampler.
Pressing No aborts the process.
19) If used, remove the MAS-100 Family Calibration Disk or the Petri dish.
20) This procedure may be repeated for all other sampling heads.
21) End of process.