User Manual MAS-100 Regulus en.docx, Version: 4.0
12) After successful login, the PC Software starts in the Adjustment/Calibration menu.
13) Select the perforated lid of the sampling head:
300×0.6 or 400×0.7
14) Choose the calibration validity:
Values from 0 to 12months may be entered.
A value of 0 disables the calibration validity
check upon starting up the air sampler.
15) Activate the optional function Measurement block-
ing upon expired calibration, if desired.
16) The absolute Maximum allowable deviation is set at
2.5% for this procedure and cannot be modified.
Note
The standard factory setting for the Calibration validity is 12months.
Note
The standard factory setting for the absolute Maximum allowable deviation is
5%.
Note
Since PC software V3.1.14, the Disable calibration validity function is only ac-
tive for Factory user level.