Page 153 of 188
User Manual MAS-100 Regulus en.docx, Version: 4.0
12) After successful login, the PC Software starts in the Adjustment/Calibration menu.
13) Select the perforated lid of the sampling head:
300 × 0.6 or 400 × 0.7
14) Choose the calibration validity:
– Values from 0 to 12 months may be entered.
– A value of 0 disables the calibration validity
check upon starting up the air sampler.
15) Activate the optional function Measurement block-
ing upon expired calibration, if desired.
16) The absolute Maximum allowable deviation is set at
2.5 % for this procedure and cannot be modified.
Note
The standard factory setting for the Calibration validity is 12 months.
Note
The standard factory setting for the absolute Maximum allowable deviation is
5%.
Note
Since PC software V3.1.14, the Disable calibration validity function is only ac-
tive for Factory user level.