The maximum allowed Deviation from target air flow is ±4.0 %.
The air sampler must be re-adjusted according to chapter 7.5, chapter 7.4, chapter 8.4
or chapter 8.5 if the calibration result is outside the specified range.
17) To create a calibration certificate and print the displayed data afterwards, press on
the hyperlink Go to the as left certificate at the bottom of the result page.
18) A new tab will open in the brower.
Press Print this page to either print the certificate on a printer or virtual PDF printer.