14) Set the adjustment and calibration parameters:
Perforated lid type: 300 × 0.6 or 400 × 0.7
Run time check
(4)
: ON / OFF
Calibration reminder: ON / OFF
Calibration validity: 1 – 12 months
Activate calibration validity immediately: ON / OFF
Block measurements if calibration is expired
(5)
: ON / OFF
15) Press the Start button to initiate the procedure, or
Press the Stop button to abort the process.
16) Once the process is finished the Website will be updated with the result:
(4)
If activated, this feature compares the air samplers internal CPU clock with the PC’s watch. An error will be generated and logged in the air samplers
error log if the deviation is out of specification.
(5)
If. activated, this feature will make it impossible to run a sampling if the calibration of the air sampler has expired.