14) After confirming a successful calibration, the air
sampler will display the key and padlock symbols.
The user has two options:
– Press the button to end the process and shut
down the air sampler.
The calibration will be activated after restarting
the air sampler.
or
– Press the ► button to activate the calibration re-
minder immediately and continue with step 15.
15) The air sampler will display the calibration validity.
Press the ► button to continue.
Note
The calibration validity can only be modified via the Browser-based software.
16) Press the button to get to the Exit ? option.
17) Press the ► button to exit.
18) The air sampler is in the main menu.
19) Remove the MAS-100 Family Calibration Disk or the Petri dish.
20) End of process.