User Manual MAS-100 Regulus en.docx, Version: 4.0
14) After confirming a successful calibration, the air
sampler will display the key and padlock symbols.
The user has two options:
Press the button to end the process and shut
down the air sampler.
The calibration will be activated after restarting
the air sampler.
or
Press the button to activate the calibration re-
minder immediately and continue with step 15.
15) The air sampler will display the calibration validity.
Press the button to continue.
Note
The calibration validity can only be modified via the Browser-based software.
16) Press the button to get to the Exit ? option.
17) Press the button to exit.
18) The air sampler is in the main menu.
19) Remove the MAS-100 Family Calibration Disk or the Petri dish.
20) End of process.