12) After successful login, the PC Software starts in the Adjustment/Calibration menu.
13) Select the perforated lid of the sampling head:
300 × 0.6 or 400 × 0.7
14) The absolute Maximum allowable deviation is set at
2.5 % for this procedure and cannot be modified.
15) Start the process with the button.
Pressing afterwards aborts the process.
16) Once the process is done, the final calibration re-
sult will be displayed by the value Deviation from
the Target Air Flow.
A calibration was successful if:
Deviation from the Target Air Flow is within (±) the
Maximum allowed deviation.
Successful calibration, value
marked green:
Failed calibration, value marked
red: