Page 91 of 188
User Manual MAS-100 Regulus en.docx, Version: 4.0
Note
The calibration also fails if at least one of the values in the row of Devia-
tion from target air flow is out of the allowable range.
17) Press Create test protocol.
18) Enter the Test protocol Nr.,
then Press OK.
19) The test protocol is exported automatically to a PDF file in a separate window.
The PDF file can be printed and signed if required.
20) If used, remove the MAS-100 Family Calibration Disk or the Petri dish.
21) End of process.