If patch has connectivity issues the skin can be lightly abraded at each electrode site
with a rough washcloth, or an abrasive pad.
Step 3 - Electrode Patch Application
Applying Tele
Patch Patch:
Prepare skin, as shown in Step 2.
When ready to position Electrode Patch on skin, remove backing from Electrode Patch,
apply to chest, using placement in the illustration. This can be done before inserting
Pendant in the cradle or after.
Press Electrode Patch firmly onto skin with fingers once correctly positioned on the left
side of chest (Below collarbone).
Once it is firmly adhered to skin peel remaining white tabs on the patch so the whole
surface of the patch is adhered to skin.
THE MEDICOMP ECG ELECTRODE PATCH IS SINGLE USE, NON-STERILE, AND DISPOSABLE
AND IS TO BE USED ON INTACT (UNINJURED) SKIN.
PRODUCT SHELF LIFE IN INTACT, ORIGINALLY SEALED PACKAGING IS 12 MONTHS.
ALLOW 15 MINUTES FOR SKIN TO ABSORB THE ELECTRODE GEL, WHICH MAY HELP
PROCEDURE INITIALIZATION ERRORS FROM OCCURRING.
CONDUCTIVE PARTS OF ELECTRODES AND ASSOCIATED CONNECTORS FOR TYPE BF OR
CF APPLIED PARTS, INCLUDING NEUTRAL ELECTRODE, SHOULD NOT CONTACT OTHER
CONDUCTIVE PARTS, INCLUDING EARTH.