Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
30 A/m (Both 50Hz, 60Hz)
NOTE: EUT PM750 Pendant shuts down between +/-8 and 15kV Air Discharge. It is recommended that you
restart the device (wait 30 seconds after power down) upon occurrence of this event.
GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY
The PM750 is intended for use in the electromagnetic environment specified below. The customer or user of
the PM750 should assure that it is used in such an environment.
Immunity Standards per 60601-1-2 :2014, (Home Healthcare Environmental) and ETSI EN301 489-1 V1.9.2
(2011-09)
Immunity Test IEC 60601
Compliance Level
Conducted RF IEC
61000-4-6
3Vrms 150kHz to 80MHz,
80% AM 1kHz
3Vrms 150kHz to 80MHz,
80% AM 1kHz
3Vrms 150kHz to 80MHz,
80% AM 1kHz
Radio Frequency
Electromagnetic Field
Amplitude 61000-4-3
10 V/m, 80 MHz to 2.7
GHz, 80% AM 1kHz
10 V/m, 80 MHz to 2.7
GHz, 80% AM 1kHz
10 V/m, 80 MHz to 2.7
GHz, 80% AM 1kHz
Radiated RF IEC
61000-4-3
385MHz: 27V/m @ 18Hz
pulse modulation
450MHz: 28V/m @ FM
modulation
9V/m @
217 Hz pulse modulation
810MHz, 870MHz,
930MHz: 28V/m @
18Hz Pulse modulation
1720MHz, 1845MHz,
1970MHz: 28V/m @
217 Hz Pulse Modulation
pulse modulation,
450MHz: 28V/m @ FM
modulation
710MHz, 745MH,
780MHz: 9V/m @ 217 Hz
pulse modulation
810MHz, 870MHz,
930MHz: 28V/m @
18Hz Pulse modulation
1720MHz, 1845MHz,
1970MHz: 28V/m @
217 Hz Pulse Modulation
385MHz: 27V/m @ 18Hz
pulse modulation,
450MHz: 28V/m @ FM
modulation
9V/m @ 217 Hz pulse
modulation
810MHz, 870MHz,
930MHz: 28V/m @
18Hz Pulse modulation
1720MHz, 1845MHz,
1970MHz: 28V/m @
217 Hz Pulse Modulation
Tele
Patch™ Cardiac Monitor PM750
USER MANUAL - UTM0000701-05E
| APRIL 4, 2019