Pulse modulation
5240MHz, 5500MHz,
5785MHz: 9V/m @
217 Hz Pulse modulation
217Hz Pulse modulation
5240MHz, 5500MHz,
5785MHz: 9V/m @
217 Hz Pulse modulation
2450MHz: 28V/m @ 217Hz
Pulse modulation
5240MHz, 5500MHz,
5785MHz: 9V/m @
217 Hz Pulse modulation
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND THE PM750
The PM750 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of PM750 can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the PM750
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
of transmitter (W)
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.167ҀP
1.167ҀP
2.333ҀP
0.01 .12 .12 .23
0.1 .37 .37 .74
1 1.2 1.2 2.3
10 3.7 3.7 7.4
100 12 12 23
For transmitters rated at a maximum output power not listed above the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE: 1 - At 80 MHZ and 800 MHZ, the separation distance for the higher frequency range applies.
NOTE: 2 - These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Tele
Patch™ Cardiac Monitor PM750
USER MANUAL - UTM0000701-05E
| APRIL 4, 2019
79