• For additional troubleshooting instructions, call Medicomp’s Patient Support: 877-996-5553.
LOW VOLTAGE IN CHANNEL B
(BLACK [B-], RED [A+] ELECTRODE PAIR
The Handset's message for a low voltage in Channel B is: Low B, Move Black
First, move the black electrode
Low Voltage in Channel B, Move Black After repositioning,
Press 'Retry'
If the Handset continues to display the error, try an electrode location on the opposite side of
the Black (B-) electrode. Don't try more than a few different locations.
• For additional troubleshooting instructions, call Medicomp’s Patient Support: 877-996-5553.
HIGH VOLTAGE IN CHANNEL A
(RED [A+]/WHITE [A-] ELECTRODE PAIR)
The Handset's message for this error is: High A, Move Red
First, move the red electrode
High Voltage in Channel A, Move Red After repositioning,
Press 'Retry'
If the Handset continues to display the error, try an electrode location on the opposite side of
the Red (A+/B+) electrode. Don't try more than a few different locations.
• For additional troubleshooting instructions, call Medicomp’s Patient Support: 877-996-5553.
HIGH VOLTAGE IN CHANNEL B
(BLACK [B-]/RED [A+] ELECTRODE PAIR)
The Handset's message for a high voltage in Channel B is: High B, Move Black
First, move the black electrode
High Voltage in Channel B, Move Black After repositioning,
Press 'Retry'
If the Handset continues to display the error, try an electrode location on the opposite side of
the Black (B-) electrode pictured. Don't try more than a few different locations.
• For additional troubleshooting instructions, call Medicomp’s Patient Support: 877-996-5553.