RF emissions CISPR 11 Class B The PM750 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes
Harmonic emissions IEC
61000-3-2
Not Applicable
Voltage Fluctuations/ flicker
emissions IEC 61000-3-3
Not Applicable
GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY
The PM750 is intended for use in the electromagnetic environment specified below. The customer or user of
the PM750 should assure that it is used in such an environment.
Immunity Standards per 60601-1-2 :2014, (Home Healthcare Environmental) and ETSI EN301 489-1 V1.9.2
(2011-09)
Immunity Test
IEC 60601
Test Level/Limits
Compliance Limits Compliance Level
NOTE: EUT PM750 Pendant
shuts down between +/-8
and 15kV Air Discharge. It is
recommended that you
restart the device (wait 30
seconds after power down)
upon occurrence of this
event.
Electrostatic
discharge (ESD) IEC
61000-4-2
Contact 2kV,4kV,8kV, 15kV
2kV,4kV,6kV 8kV (+/-)
Contact 2kV,4kV,8kV,
15kV (+/-) AIR
2kV,4kV,6kV 8kV (+/-)
Contact 2kV,4kV,8kV, 15kV
(+/-) AIR
Electrical fast
transient/burst IEC
61000-4-4
±1kV Signal and Control
±2.0kV AC/DC Mains
±1kV Signal and Control
Lines
±2.0kV AC/DC Mains
±1kV Signal and Control
Lines
Voltage Dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
70% of Vnom, 25 cycles at
0deg
0% of Vnom, 0.5cycles at
0deg, 45deg, 90deg,
180deg, 225deg, 270deg,
315deg
0% of Vnom, 1.0 cycles at
70% of Vnom, 25 cycles at
0deg
0% of Vnom, 0.5cycles at
0deg, 45deg, 90deg,
180deg, 225deg, 270deg,
315deg
70% of Vnom, 25 cycles at
0deg
0% of Vnom, 0.5cycles at
0deg, 45deg, 90deg, 180deg,
225deg, 270deg, 315deg
0% of Vnom, 1.0 cycles at
Tele
Patch™ Cardiac Monitor PM750
USER MANUAL - UTM0000701-05E