recording
Early low main battery
warning
Orange LED blink 1 per
second
Beep 1 second duration every
4 minutes
Late low main battery
warning
Orange LED blink 3 times per
second
Beep 1 second duration every
30 seconds
TROUBLESHOOTING
Setup Details and Troubleshooting Overview
This section helps resolve any problems that may be encountered while you are setting up the
Tele
Patch. Reduce the likelihood of seeing errors by making sure fresh Electrode Patch or fresh
individual electrodes are used, as appropriate, the cable is in good condition and securely
connected to the Cable Cradle, and care has been taken to perform the recommended skin
preparation.
Some problems are unavoidable due to a patient's particular heart condition or the unexpected
failure of a component. However, these cases are infrequently encountered.
When the Tele
Patch is connected and begins its startup sequence, it assesses the quality of the
ECG signal. The TelePatch system will signal if there is a problem with the connections or if the
signal voltage is too high or too low for ideal auto-capturing. Messages on the Handset's display
will direct any necessary corrections.
Information is also provided in this section to assist with the setup procedure of any
replacement or exchanged Tele
Patch system component (e.g. Handset or Pendant). Replaced
components of the TelePatch will follow a slightly different setup procedure than the standard
method.
TelePatch System Hookup Testing Information
Quality Hookup Tests Overview
The Tele
Patch Quality Hookup tests evaluate cable connections, skin preparation, and
signal-voltage conditions. This is a test that is performed automatically by the TelePatch
System, with no indication to, or participation from, the user or patient until specifically
directed by specific commands on the Handset.
The Quality Hookup test starts once either:
Tele
Patch Cardiac Monitor PM750
USER MANUAL - UTM0000701-05E
| APRIL 4, 2019
63