Have a question or need assistance?
Call Medicomp’s Patient Support: 877-996-5553
Use a tissue, lightly moistened with alcohol or a hospital-grade cleanser such as Sporicidin –
(Do not saturate) - wipe off debris, dirt or other matter that may gather on either the Handset
or Pendant of the Tele
Patch System. If the Pendant has been dropped in a basin of water (i.e.:
sink or toilet) and will not power back on, contact Customer Support for a replacement.
VISUAL AND AUDIBLE INDICATORS
Normal Performance Indicators
Pendant Action Visual Indicator Audible Indicator
Pendant power up Green, blue, and orange lit
simultaneously for 1.5 seconds.
Beep High-Low-High for 0.5
second per frequency while
LEDs are lit.
Pendant ready to start
procedure
Green LED blinks fast None
Procedure initiated: Press
Symptom button 3 times
(Holter Only)
Green LED solid for 1.5 seconds
Beep High-Low-High for 0.5
second duration for each
beep
Procedure initiated not in
cradle: Press Symptom
button 3 times (Holter
procedure only)
Orange and blue blinks fast in
unison
Beep 4 times in unison with
LEDs
Procedure in progress: not in
cradle
3 LED sections: continuous
flashing
None
Pendant removed from
Cradle
3 LED sections: continuous
flashing
None
Pendant placed in cradle with
procedure running BT
OFF/Holter
Green LED: every 4 seconds Beep for 2 seconds before
entering this mode for the
first time
Symptom Button Press on
cables or patch: Record
Symptom
Green LED lit for 1 second Beep for 1 second
Tele
Patch™ Cardiac Monitor PM750
USER MANUAL - UTM0000701-05E
| APRIL 4, 2019
61