Have a question or need assistance?
Call Medicomp’s Patient Support: 877-996-5553
The Medicomp TelePatch Handset
Monitoring Screen
●
The Tele
Patch procedure is now running.
IT IS VERY IMPORTANT TO TURN THE PENDANT OFF BEFORE CHANGING BATTERY,
OTHERWISE DATA MAY BE LOST. THE POWER BUTTON IS LOCATED ON THE FACE OF THE
PENDANT.
DEPENDING ON THE SETTINGS, AND THE PATIENT PREPARATION, SOME INTERMITTENT
CONFIRMATION AND/OR ERROR SCREENS MAY APPEAR DURING THE INITIALIZATION.
PLEASE PERIODICALLY CHECK THE PENDANT AND/OR Handset TO ENSURE THEY ARE
BOTH IN MONITORING MODE. THIS MEANS THE PENDANT IS BLINKING BLUE LED LIGHTS,
AND THE Handset DISPLAY STATES ‘MONITORING’.
THE TELEPATCH SYSTEM IS DESIGNED TO OPERATE IN ENVIRONMENTS THAT MAY
EXPERIENCE TRANSIENT ELECTRICAL EVENTS LIKE ELECTROSTATIC DISCHARGE (ESD).
TELEPATCH IS DESIGNED TO BE RESISTANT TO SUCH EVENTS AND WILL TYPICALLY CONTINUE
TO OPERATE NORMALLY. A SEVERE ESD EVENT MAY CAUSE THE DEVICE TO SHUT DOWN IN
ORDER TO PROTECT THE MEMORY. IF A SHUTDOWN OCCURS, MANUALLY POWER THE DEVICE
ON USING THE POWER BUTTON ON THE FRONT OF THE DEVICE. ONCE POWERED ON,
TELEPATCH SHALL RESUME NORMAL OPERATION WITH NO LOSS OF FUNCTIONALITY OR
INFORMATION.
Best Practices for using the Handset and the Pendant:
Tele
Patch™ Cardiac Monitor PM750
USER MANUAL - UTM0000701-05E
| APRIL 4, 2019
44