Environmental condions
Operaon Temperature: 15 to 40°C
Relave humidity: 5% to 95% (not condensing)
Storage Short-term/transport: Temperature: -20 to +50°C
Relave humidity: 5 to 95%
(not condensing)
Long-term: Temperature: Room temperature
(about 20°C)
Relave humidity: 20 to 80%
(not condensing)
Descripon of the modes
CPAP Standard CPAP; addional funcon: Leak-Assist
Apnea CPAP CPAP with apnea detecon (automacally generated pushes);
addional funcons: Backup (backup rate: 5 to 120 L/min as a
funcon of T
insp.
(minimum 0.2 s), Leak-Assist
NIPPV Bi-level CPAP with measurement of the I:E rao as a funcon of
the set rate and inspiraon me (minimal I:E rao 1:1.5), addional
funcon: Leak-Assist
High Flow Flow emission for Nuow® nasal cannulas in ow range of 0 to 12
L/min, adjustable O
2
ush
Standby No ow, deacvated alarms, sengs made are saved, keeps
medin-NC3® ready for use
Soware sengs
Languages EN, DE, FR, IT, NL, ES, EL, CS, PL, NO, DA, SV, RU, LT, ZH, TR, JA, RO
Pressure units mbar or cmH
2
O
Pressure scale Three sengs: 0 to 10, 0 to 15 and 0 to 20 mbar
Respiraon rate Oponally shown in display
FiO
2
Oxygen concentraon
Seng range: 21 to 100% (in 1% increments)
Measurement range: 21 to 100%
Alarm sengs: ± 2-5% dierence
O
2
ush (level: +5, +10, +15% (vol.) above set target oxygen
concentraon) for 60 s
Accuracy:
± 5% (2 to 4 L/min),
± 2% (from 4 L/min)
Alarms Visual (LED & display) and acousc (adjustable push alarm sengs)
Alarm countdown
Safety Mechanical overpressure valve (opening pressure 6 kPa (= 60 mbar))
External data USB/RS232 port, export of live and trend data
Operang me The medin-NC3® can be used for connuous, long-term operaon
up to 4 weeks without a restart in the interim.
medin Medical Innovaons GmbH
Adam-Geisler-Str. 1
82140 Olching, Germany
+49 8142 448 460
info@medin-medical.com
www.medin-medical.com
IFU_medin-NC3 EN
Rev 05 Status as of 28.04.2021
10 of 68