9 Cleaning and Maintenance
Warning:
In order to avoid res, oxygen must not come into contact with oil or grease. It is therefore essenal to ensure that
all parts of the medin-NC3® and the CPAP system which come into contact with oxygen (e.g. gas feed tubes, O
2
inlet to the medin-NC3®, paent outlet, paent tubes and Medijet®) remain free from oil and grease even during
cleaning, repairs and maintenance.
While the device is in operaon on a paent, no cleaning, maintenance or repairs may be performed on the
medin-NC3® – including all points listed in chapter 9.
medin oers two dierent warranty terms:
- 3 years (REF 30-912) from the date of producon
- 5 years (REF 30-913) from the date of producon
These warranty expansions only relate to components with defects caused by producon, wear parts, as well as
soware problems. In addion, the customer must demonstrate that all required maintenance was performed.
Please note:
The following are not included in the warranty: physical damage, environmental damage (e.g. water or grease in
the device connecon, excessively high voltage), improper use (e.g. excessively high gas feed pressure) as well as
consumables (e.g. oxygen cells, lters, baeries).
9.1 Cleaning
The surface of the medin-NC3® should be disinfected using a slightly damp cloth
1
before rst use and every me
aer it has been used on a paent. 70% isopropyl alcohol soluon can be used for this purpose.
1
Slightly damp = Take a wet cloth and wring it out. Then wrap a dry cloth around the wet cloth and wring out
both of these cloths together once again.
Warning:
- The device must never be sterilized or immersed in a liquid soluon.
- No abrasive substances may be applied to the surface.
- No liquid may penetrate the device.
9.2 Device maintenance
The medin-NC3® must undergo maintenance and a funcon check in accordance with the manufacturer's
instrucons every 12 months. This must be carried out by trained specialists using appropriate measuring and
tesng equipment.
The following acons must be performed during this maintenance:
- The oxygen cells are regenerated or their service life is checked.
◦
Type MLF-16 (REF MLF-16) oxygen cells must be replaced if they have less than one
year of service life le or no later than aer three years.
- Test of the device’s funconality and sensor calibraon.
- The internal 3-volt baery must be replaced as required or every three years at the latest. The
replacement baery must be a CR123A powerone from VARTA Microbaery GmbH (REF 39-102).
- The internal rechargeable baery must be replaced as required or every three years at the latest. The
replacement baery must be a rechargeable baery of the same type (can be ordered from medin,
REF 39-101).
Aenon: Use only approved cleaning agents for cleaning and disinfecon.