5.8 Apnea CPAP mode
In apnea CPAP, the basic CPAP of the CPAP mode is complemented by apnea monitoring. The apnea monitoring
system assesses uctuaons in the CPAP pressure course which are caused by the paent’s respiraon and responds
to phases in which the paent does not breathe by automacally triggered pushes.
In addion to the seng, acon and informaon opons which are present in CPAP mode, such as:
- Flow (chapter 5.3)
- FiO
2
oxygen concentraon (chapter 5.3) and start and disconnuaon of the O
2
ush (chapter 5.4)
- F
insp
addional push ow and manual PUSH (chapter 5.5)
- Alarm sengs (chapter 5.13)
- Trend graphs (chapter 5.14)
there are special sengs for the apnea mode. These apnea sengs can be opened by pressing the p buon in
Figure 5-2, which oers the possibility of changing six parameters. To change a parameter, press the corresponding
buon to select it, turn the dial to the desired seng, and press the dial to conrm your selecon.