4 First Use
Warning:
If errors are detected when the medin-NC3® is checked before rst use or if the medin-NC3® is found to be damaged
or does not behave as expected or as intended in the user manual, it must not be connected to a paent under
any circumstances.
The power supply unit should not be used if it has visible damage to the housing or cord.
The power supply unit should not be used in the direct vicinity of the paent since it may heat up to high
temperatures.
4.1 CPAP system components
A CPAP system based on a medin-NC3® comprises the following components:
- CPAP driver: the medin-NC3®
- Medijet® nCPAP generator together with a suitable prong or mask
- An acve respiratory gas humidier to humidify the respiratory gas (recommended)
- A paent tube circuit set that ts the paent gas outlet, CPAP pressure meter inlet of the medin-
NC3®, the Medijet® and the respiratory gas humidier selected
- A bonnet of the correct size
- A system for monitoring the paent’s oxygen saturaon and checking the ecacy of CPAP therapy
(this is an essenal condion for the use of the medin-NC3®)
Please note:
The medin-NC3® should always be used in combinaon with an acve respiratory gas humidier in order to avoid
the paent’s airways drying out and cooling.
The CPAP system must be established and the device checked before starng any CPAP therapy using the
medin-NC3®. This check starts automacally whenever the device is started up.
The device should not be connected to a paent during set-up and system start. Moreover, during set-up, system
start and operaon, the user should not simultaneously touch the paent and the device. This applies above all to
the power supply, the RS232 or USB interface and the accessible metal parts of the pneumac unit (gas connecons,
overpressure valve and mounng screws).