Guidance and manufacturer’s declaraon – electromagnec emissions
The medin-NC3® is intended for use in the electromagnec environment specied below. The customer or user
of the medin-NC3®, or the organizaon responsible, should assure that it is used in such an environment.
Emissions test Compliance
Electromagnec environment – guidance
RF emissions according to CISPR11 Group 1
The medin-NC3® uses RF energy only for its
internal funcon. Therefore, its RF emissions are
very low and not likely to cause any interference
in nearby electronic equipment.
RF emissions according to CISPR11 Class A
The medin-NC3® is intended for use in professional
healthcare instuons (CISPR11, group 1, class A).
Harmonic emissions according to
IEC 61000-3-2
Class A
Voltage uctuaons/icker
emissions according to
IEC 61000-3-3
Complies
medin Medical Innovaons GmbH
Adam-Geisler-Str. 1
82140 Olching, Germany
+49 8142 448 460
info@medin-medical.com
www.medin-medical.com
IFU_medin-NC3 EN
Rev 05 Status as of 28.04.2021
60 of 68
10 Electromagnec Compability
The informaon in this secon is provided in order to enable the operator of the medin-NC3® to decide whether
the medin-NC3® is suitable for its electromagnec environment.
Aenon:
- The medin-NC3® is a medical electrical device. Consequently, in order to guarantee the funcon of
the medin-NC3®, precauons with regard to electromagnec compability must be taken and the
medin-NC3® must be set up and used in accordance with the condions set out below.
- The characteriscs of the medin-NC3® determined by transmissions permit it to be used in industrial
sengs and hospitals (CISPR 11, class A). When used in a residenal seng (for which class B is
usually necessary, according to CISPR 11), this device may not oer appropriate protecon from
radiocommunicaon services. If necessary, the user must take correcve measures such as moving
or reorienng the device.
- Portable or mobile RF communicaons equipment (e.g. mobile phones) should not be used less
than 30 cm near medin parts and cables associated with the medin-NC3®. Failure to observe these
instrucons can lead to a reducon in the performance of the device.
Please note:
Electromagnec interference may aect the device and this can be manifested through alarms (for a detailed
descripon, see chapter 8) or limitaons in normal funcon of the medin-NC3® (see chapter 8.3).
10.1 Electromagnec transmission