7.2 Components of the rest of the CPAP system
The medin-NC3® must be used in combinaon with the Medijet® nCPAP generator and the appropriate accessories
(bonnets, masks, prongs and tube circuit sets).
Warmed and humidied Nuow® nasal cannulas (High Flow therapy) can also alternavely be connected in the
corresponding (High Flow) mode.
For more specic informaon on this, visit the homepage www.medin-medical.com.
Warning:
The medin-NC3® can only be operated in combinaon with the Medijet® nCPAP generator and the appropriate
tube circuit sets, bonnets, masks and prongs. If used with any other nCPAP generator, the funconality of the
device cannot be ensured, the correct alarm funcon cannot be guaranteed, sucient CPAP pressure will not be
generated and apnea detecon will not work.
The correct mode must be selected for the accessory connected in each case. Nasal cannulas may not be used in
CPAP mode and Apnea CPAP mode and the Medijet® may not be operated in High Flow mode.