Error message text Possible cause Correcon
Connect power supply unit –
baery low
The internal baery is almost
discharged.
Please connect the power supply
unit to the mains.
Fan blocked The fan is blocked or defecve. Check whether the venlaon
slits are clear or contact a service
technician.
Gas feed pressure low,
check O
2
The oxygen feed pressure is too low
(< 3 bar).
Please check oxygen supply (tube
connected?).
(1)
Gas feed pressure high,
check O
2
The oxygen feed pressure is too high
(> 7 bar).
Please check oxygen feed.
(1)
Check gas feed air The blower reported an error. Clean blower lter or contact a
service technician.
Oxygen calibraon failed,
check O
2
The oxygen calibraon could not be
performed correctly. The medin-NC3®
cannot be operated without oxygen.
Please check whether the
oxygen cell is working properly
and whether the medin-NC3® is
correctly connected to the oxygen
feed.
Device error – device
shutdown necessary
There is a defect in the device’s
pneumac system, at the overpressure
valve or rechargeable baery. The
medin-NC3® can no longer be used and
must be switched o.
Check overpressure valve
(blocked?) or contact a service
technician.
Error during tube test –
restart
An error occurred during the tube test
(e.g. leakage).
Check impermeability and repeat
the tube test.
Device error – paent
pressure. Device must be
replaced
The paent pressure sensor is causing
an error. The device cannot be used and
must be replaced
Please contact a service
technician.
3V baery empty – baery
has to be exchanged
The internal 3V baery is empty and
must be replaced.
Please contact a service
technician.
Baery needs replacing –
device funcon in baery
mode cannot be guaranteed.
Device must be connected to
mains for use!
The internal/external rechargeable
baery is too old or was used too
frequently and must be replaced.
Do not use the medin-NC3® on
rechargeable baery power and
contact a service technician.
Maintenance necessary!
Next maintenance necessary:
YY.MM.DD
The last maintenance of the
medin-NC3® took place at least 11
months ago and must be repeated no
later than the date indicated.
Please contact a service
technician.
Blower lter blocked The blower lter of the device is
clogged and must be cleaned or
replaced.
Please clean blower lter
medin Medical Innovaons GmbH
Adam-Geisler-Str. 1
82140 Olching, Germany
+49 8142 448 460
info@medin-medical.com
www.medin-medical.com
IFU_medin-NC3 EN
Rev 05 Status as of 28.04.2021
49 of 68
Table 3: Error messages at system start
(1)
If this error should occur more frequently despite correctly connected tubes, please contact the person
within your organizaon who is responsible for the central gas supply.
8 Alarms and Error Messages
8.1 Error messages during system start-up