59 of 68
9.6 Filter
The medin-NC3® contains two replaceable lters: a sucon lter of the blower as well as a coarse lter to avoid
soiling by the device fan.
The blower lter must be cleaned every 2 months by tapping or knocking (dry). The lter needs to be replaced
with a new one aer 12 months.
To replace the lter, the holder must be ipped open to the side and the lter cassee must be taken out. The
lter can now be cleaned. Then the lter cassee including the lter can be inserted once again and the holder
can be locked.
Figure 9-1: Fan lter / blower lter
Please note:
Only the replacement lter (REF 39-317 for fan lters, REF 30-203 for blower lters) may be used.
The coarse lter must also be cleaned every 2 months by tapping or knocking (dry). The lter should be replaced
with a new one aer 12 months.
To replace the lter, it is pulled out from above and reinserted aer cleaning.
Please note:
Only a type PPI 30 lter should be used.
Warning:
To avoid an electrical shock, the housing of the power supply unit of the medin-NC3® should not be opened.
Modicaon of the power supply unit is not permied (terminaon of the warranty).
Maintenance/repairs/modicaons:
The power supply unit FSP090-RACM1 of the medin-NC3® is maintenance-free. It should not be opened.
The power supply unit may only be repaired by authorized specialists.