9.6 Filter
The medin-NC3® contains two replaceable lters: a sucon lter of the blower as well as a coarse lter to avoid
soiling by the device fan.
The blower lter must be cleaned every 2 months by tapping or knocking (dry). The lter needs to be replaced
with a new one aer 12 months.
To replace the lter, the holder must be ipped open to the side and the lter cassee must be taken out. The
lter can now be cleaned. Then the lter cassee including the lter can be inserted once again and the holder
can be locked.
Figure 9-1: Fan lter / blower lter
Please note:
Only the replacement lter (REF 39-317 for fan lters, REF 30-203 for blower lters) may be used.
The coarse lter must also be cleaned every 2 months by tapping or knocking (dry). The lter should be replaced
with a new one aer 12 months.
To replace the lter, it is pulled out from above and reinserted aer cleaning.
Please note:
Only a type PPI 30 lter should be used.
Warning:
To avoid an electrical shock, the housing of the power supply unit of the medin-NC3® should not be opened.
Modicaon of the power supply unit is not permied (terminaon of the warranty).
Maintenance/repairs/modicaons:
The power supply unit FSP090-RACM1 of the medin-NC3® is maintenance-free. It should not be opened.
The power supply unit may only be repaired by authorized specialists.