3.2 Power supply
3.2.1. Mains operaon
The power supply unit (REF 30-113) of the medin-NC3® may only be connected to a power supply with the
following properes:
Voltage: 100 – 240 V~ (alternang current)
Frequency: 50 – 60 Hz
Strength of current: at least 1.1 A
The power supply unit of the medin-NC3® has a C7 appliance inlet for connecon to the mains grid. The power line
between the power supply unit of the medin-NC3® and the mains grid must therefore meet the following condions:
Power supply side: plug appropriate for the country
Device side: C8 plug in accordance with EN 60320
Minimum rated voltage: the minimum rated voltage of the power line must be at least as high as the
voltage of the local mains supply (recommended: 240 V)
Minimum rated current: 1.1 A
Maximum length 2 meters (in order to achieve EMC in accordance with DIN EN 60601-1-2)
Spare power lines with EU plugs can be ordered under REF 39-115 (supplied as standard as part of the power
supply unit). Alternave power lines for other countries are available under these numbers:
Aenon:
The medin-NC3® may only be run on the local mains grid using the power supply unit provided by the manufacturer
(REF 30-113) and suitable power lines, otherwise the funconality and EMC of the device cannot be guaranteed.
The device is disconnected from the mains by pulling the power supply unit cable of the medin-NC3® out of the
electrical outlet. Thus when assembling the medin-NC3®, it should be ensured that this plug is always freely
accessible and can be disconnected from the mains without diculty. The medin-NC3® does not contain any
switch which disconnects it from the mains.
3.2.2. Baery operaon
The medin-NC3® contains an internal as well as an external baery and can work on a fully charged baery for
up to four hours.