Soware version Hardware changes Descripon
SW version V 1.01 /
A.12.00 together with
display macro version
V101
First series version First version
SW version V 1.10 /
A.14.00 together with
display macro version
V110
First series version Change in the shutdown behavior: mulstage
query
Faster calculaon of the mean pressure
values in NIPPV
Opmizaon of the Leak-Assist and change
in the minimal total ow from 3 L/min to
4 L/min
Reading the pneumac self-test in the service
menu
The message “Device error – device
shutdown necessary” is triggered more
rarely; instead, the message “Press alarm
reset to restart ow” is iniated. This change
within the emergency program enables more
connuous applicaon of the nCPAP therapy
with the medin-NC3®.
White informaon signal now blinks
when the baery is being charged and is
connuously lit when the power supply unit
is connected
Pneumac system Oxygen valve 2871 replaced with a variant
with idencal construcon
External baery compartment Hand knurl replaced by rounded head screw
Blower cable Longer: 300 mm instead of 200 mm
DC port New variant
Pneumac system Measuring aperture replaced by a laminar
ow element, addional diode as a protecve
funcon
Packaging label Addion of LOT
Reinforcing bracket for housing Installaon of reinforcing bracket on hood
Fan Fan MFB60D05A replaced by
PLB60B05FSE35A
Glass front More adherent adhesive tape
medin Medical Innovaons GmbH
Adam-Geisler-Str. 1
82140 Olching, Germany
+49 8142 448 460
info@medin-medical.com
www.medin-medical.com
IFU_medin-NC3 EN
Rev 05 Status as of 28.04.2021
68 of 68
14 History of the Soware Versions and Hardware Changes
The soware of the medin-NC3® consists of various soware packets: part of them determines the operang
properes of the medin-NC3®, the alarms and the funconal range. The display macro version establishes the
possible display languages. In doing so, each SW version is dependent on a parcular version of the display macro.