Guidance and manufacturer’s declaraon – electromagnec immunity
The medin-NC3® is intended for use in the electromagnec environment specied below. The customer or user of
the medin-NC3® or the organizaon responsible should assure that it is used in such an environment.
Immunity tests
IEC 60601
test level
Compliance level Electromagnec environment – guidance
Electrostac discharge
(ESD) according to
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
Floors should be wood, concrete or ceramic
le. If oors are covered with synthec
material, the relave humidity should be at
least 30%.
Electrical fast
transients/bursts
according to
IEC 61000-4-4
± 2 kV for
power supply
lines 100 kHz
repeon
frequency
± 1 kV for
input/output
lines 100 kHz
repeon
frequency
± 2 kV for power
supply lines
100 kHz
repeon
frequency
± 1 kV for
input/output
lines 100 kHz
repeon
frequency
Mains power quality should be that of a typical
commercial or hospital environment.
Surges according to
IEC 61000-4-5
± 0.5 kV, ± 1 kV
dierenal mode
± 0.5 kV, ± 1 kV,
± 2 kV common
mode
± 0.5 kV, ± 1 kV
dierenal mode
± 0.5 kV, ± 1 kV,
± 2 kV common
mode
Mains power quality should be that of a typical
commercial or hospital environment.
medin Medical Innovaons GmbH
Adam-Geisler-Str. 1
82140 Olching, Germany
+49 8142 448 460
info@medin-medical.com
www.medin-medical.com
IFU_medin-NC3 EN
Rev 05 Status as of 28.04.2021
61 of 68
10.2 Electromagnec immunity