Note:
As long as an O
2
ush is acve, the target oxygen concentraon FiO
2
cannot be changed. Only aer a manual
premature end or aer an automac end of the O
2
ush can the FiO
2
seng be changed once again.
5.5 Push (not in High Flow mode)
In addion to the basic CPAP mode, the medin-NC3® can increase the CPAP pressure if necessary for a short me.
These periods are called "pushes" and can be used to smulate the paent. A push is determined by the addional
ow used for it and its duraon.
Addional ow during a push - F
insp
:
In eld c in Figure 5-2, set an addional ow (F
insp
) by pressing the buon to acvate the seng and turning the
dial to select a value which immediately becomes acve for the next push and then pressing the dial to conrm. If
the dial is not pushed, the change is made nonetheless. During a push, the medin-NC3® emits, as a ow, the sum
of the set basic ow (Flow) in Figure 5-2 eld a and the addional push ows (F
insp
).
As an addional push ow F
insp
a value between 0 L/min (that is, essenally no push is triggered) and a maximum
value dependent on the basic ow set can be set. The maximum value is 17.5 L/min minus the set basic ow Flow.
Warning:
A push does not generate a breath; instead it can only smulate but not venlate!
The seng F
insp
may only be changed in small increments. The eect of each increment (= CPAP pressure that
develops during a push) must be monitored by triggering a manual push or by waing for an automac push. It
should be noted that any change becomes immediately acve for the next push, even without pressing the dial.
The CPAP pressure that develops during a push depends on the leakage of the CPAP system (parcularly the
leakage from prongs or the mask). An increase in the push ow by +1 L/min can cause an increase of up to 3.5
mbar in the CPAP pressure.
Since leakage from the mask and prongs causes a decrease in the pressure during a push, this leakage must be
checked before increasing the push ow. An increase in the push ow is not a substute for leak-proof posioning
of the mask or prong!
During a push, the set basic ow (Flow) and addionally the set push ow (F
insp
) are emied such that an increase
in the basic ow also causes an increase in the CPAP pressure during a push.
When seng the addional ow during a push, the following must be ensured:
• The alarm CPAP pressure high briey stops the ow to the paent and sounds an alarm starng at a
CPAP pressure of 18 mbar; that is, if the push is set such that its pressure is very high, a slight change
in the leakage or the expiraon against push by the paent will cause the medin-NC3® to briey stop
the ow and trigger an alarm. To prevent this, the push ow must be reduced.
• If a very high overall ow is set in the event of a push, high stac pressure may develop in the tube
system. If this exceeds 60 mbar, the Feed pressure high alarm is triggered and the ow emission to
the paent is stopped. In this case, the push ow set must be reduced and the leakage at the paent
reduced. Depending on the duraon of the high supply pressure, the medin-NC3® automacally
restarts the ow aer a short me or not unl aer conrmaon from the user (if Press alarm reset
to restart ow is displayed).
• During a push, if a ow is set that is so high that it triggers the Feed pressure high alarm and
simultaneously a CPAP pressure high alarm, it is possible that the text Press alarm reset to restart
ow may also be displayed. In this case, the medin-NC3® stops the ow unl the user restarts it by
pressing the Alarm Reset buon.