- The external rechargeable baery must be replaced as required or every three years at the latest. The
replacement baery must be a rechargeable baery of the same type (can be ordered from medin,
REF 30-150).
- The blower lter (can be ordered from medin, REF 30-203) and the fan lter (can be ordered from
medin under 39-317) must be replaced every year.
- The internal tubes must be replaced as needed or every 3 years at the latest:
• REF 30-206: blue tube, connecon of the high-pressure oxygen inlet with gas mixing unit
• REF 30-207: transparent tube, connecon of gas mixing unit with outlet block
• REF 30-204: blue blower coupling, connecon of blower with gas mixing unit
- The blower holder foam (REF 30-319) must be replaced as needed or every 3 years at the latest
- Dynamic monitoring of lifespan: aer 7000 operang hours, the pneumac unit has reached the end
of its lifespan. Replace the pneumac unit.
If an error is detected during this funcon test and maintenance, the device must not be used or connected to a
paent under any circumstances.
Warning:
The two internal baeries (3V and rechargeable) may only be replaced by the baery types listed above and only by
trained, professional service personnel and in accordance with the instrucons in the service manual.
Aer each maintenance or repair of the medin-NC3®, a complete funcon test must be performed and passed
before this medin-NC3® can be used on a paent once again.
9.3 Repairs
When performing repairs and/or replacing parts, the trained professional service personnel is obligated to report
the cause(s) of the repair to the manufacturer in the form of an error code.
The current list of error codes is available in the download area of our homepage www.medin-medical.com.
Warning:
Repairs may only be made to the medin-NC3® by specially trained, professional service personnel in accordance
with the instrucons and warnings in the service manual. Aer each maintenance or repair of the medin-NC3®, a
complete funcon test must be performed and passed before this medin-NC3® can be used on a paent once again.
9.4 Device modicaons
Warning:
The medin-NC3® may not be modied without permission from the manufacturer and subsequent appropriate
examinaons and tesng to guarantee connued safe use.
9.5 Power supply unit (cleaning, maintenance, repairs and modicaons)
Cleaning:
Prior to cleaning, the power supply unit of the medin-NC3® must be disconnected from the mains.
Do not clean with chemical cleaning agents.