Recommendaon:
In the case of a brief period in standby mode (change in the applicaon such as bonnets and/or
mask/prong changes, suconing, etc.) and acvated humidier, the medin-NC3® should be restarted
before the paent is connected to the device. This will avoid the emission of hot shots to the paent.
In the case of a longer period in standby mode (pause in CPAP therapy, skin-to-skin care, etc.), the
respiratory gas humidier should be switched o.
- In the case of a longer period in standby mode and deacvated respiratory gas humidier, the
tubing system should be hung as vercally as possible so that any condensate can ow back into the
respiratory gas humidier. The Medijet® generator should be neatly covered. In addion, the 7-day
rule regarding replacement of the tubing system applies.