Determinaon of test values for electromagnec immunity of casings against high-frequency,
wireless communicaon equipment
Test frequency
MHz
Frequency band
MHz
Radio-
communicaon
service
Modulaon
Maximum
power
W
Distance
m
Immunity
test level
V/m
385 380 to 390 TETRA 400
Pulse modulaon
18 Hz
1.8 0.3 27
450 430 to 470
GMRS 460,
FRS 460
FM
± 5 kHz deviaon
1 kHz sine
2 0.3 28
710
704 to 787
LTE band 13,
17
Pulse modulaon
217 Hz
0.2 0.3 9
745
780
810
800 to 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE band 5
Pulse modulaon
18 Hz
2 0.3 28
870
930
1720
1700 to 1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE band 1, 3,
4, 25; UMTS
Pulse modulaon
217 Hz
2 0.3 28
1845
1970
2450 2400 to 2570
Bluetooth,
WLAN 802.11
b/g/n,
RFID 2450,
LTE band 7
Pulse modulaon
217 Hz
2 0.3 28
5240
5100 to 5800
WLAN 802.11
a/n
Pulse modulaon
217 Hz
0.2 0.3 9
5500
5785