7 Sterilization
33
Selecting the program
Select by rotating between the initial state and the desired program with the program selection key 'P'.
Select the sterilization program according to whether and how the }load is packed. Also note the temperature resistance
of the load.
The following table shows which program is to be selected for which load.
Sterilization programs
Universal-Program S Quick-Program S Prion-Program S
Sterilization temperature 134 °C 134 °C 134 °C
Sterilization pressure 2.1 bar 2.1 bar 2.1 bar
Sterilization time 05:30 min 03:30 min 20:30 min
Operating time*
)
13-15 min 6-8 min 28-30 min
Drying 8-9 min approx. 1 min 8-9 min
*
)
without drying and depending on load and set-up conditions (e.g. mains voltage, air pressure, feed water supply); can
take up to 2 min longer with a cold start
Use of the respective sterilization programs
Program Packaging Especially suitable for Load
Universal-Program S single wrapping Solid instruments, simple hollow bodies, dental
transmission instruments*
)
1 kg unwrapped/
270 g wrapped
Quick-Program S unwrapped only Simple solid instruments, simple hollow bodies,
dental transmission instruments*
)
1 kg unwrapped
Prion-Program S single wrapping Instruments under suspicion of carrying the
danger of infection through abnormally altered
proteins (e.g. Creutzfeldt–Jakob, BSE) dental
transmission instruments*
)
1 kg unwrapped/
270 g wrapped
*
)
Further information is provided in the certificate of suitability (enclosed).
Selecting automatic preheating
Automatic preheating is activated in delivery state.
If preheating is activated, the cold chamber is heated up to the preheating temperature of the particular program before
program start, or this temperature is held between two program runs. This reduces program times and the accretion of
condensation, thus improving drying results.
PLEASE NOTE
The steam sterilizer must remain continually activated for the automatic preheating.
MELAG recommends activating the automatic preheating function.
To alter this setting proceed as follows:
1. Select the Function menu by pressing the '+' and '-' keys
simultaneously until the display showsFunction:Last
batch number.
2. Navigate using the '+' or '-' keys until the display shows
Function: Autom. preheating.
Programm
Start-Stop
-
+
Function:
Autom. preheating