5 Installation requirements
17
5 Installation requirements
Installation location
WARNING
Failure to comply with the setup conditions can result in injuries and/or damage to the steam
sterilizer.
n The steam sterilizer should only be setup, installed and commissioned by persons authorised by MELAG.
n The steam sterilizer is not suitable for operation in explosive atmospheres.
n The steam sterilizer is conceived for use outside the patient area. The device should be located a minimum
of 1.5m radius away from the treatment area.
Property MELAquick 12+/MELAquick 12+ p
Installation surface level and horizontal
Installation location interior of a building (dry and protected from dust)
Load-bearing capacity of installation
surface
min. 25 kg plus 5 kg feed water container and 5 kg wastewater container
Floor loading (normal operation) 2.5 kN/m²
Floor loading (hydraulic pressure test) 2.6 kN/m²
Max. altitude 2000 m
Heat emission (with max. load) approx. 0.48 kW (1.74 MJ/h)
Ambient temperature 5-40 °C (ideal range 16-26 °C)
3)
Relative humidity max. 80 % at temperatures of up to 31 °C, max. 50 % at 40 °C (decreasing in
linear fashion in-between)
Steam egress can occur during operation. Do not set up the device in the immediate proximity of a smoke detector.
Maintain clearance from materials which could suffer damage from steam.
Electromagnetic environments
When assessing the Electromagnetic Compatibility (EMC) of this device, the emitted interference threshold values for
Class B devices and the stability for operation in an electromagnetic environment as described in IEC 61326-1 were
taken as the basis. The device is thus suitable for operation in all institutions and domestic settings connected to a public
mains power supply. The floor should be made of wood or concrete or be tiled with ceramic tiling. If the floor is fitted with
synthetic material, the relative humidity must amount to a minimum of 30%.
3)
Only MELAquick 12+: The installation location must be selected so that the temperature in the area of the sterile filter an along the
compressed air hose cannot exceed 40°C.