6 First steps
27
Position Description Art. no. contained in
8.2 Cooling water outlet fitting -- Vacuklav
8.3
Tap water supply hose EN 1717
ME24930 Vacuklav
9 Water stop (optional) ME01056 --
10 Water tap with safety combination (on site) -- --
Connecting compressed air (only MELAquick 12+)
NOTICE
According to
}
EN 13060, the steam sterilizer must not be operated without the sterile filter integrated
in the compressed air hose.
1. Connect a free end of the compressed air hose (Ø 6mm) to the quick coupling piece of the sterile filter pre-fitted to
the device.
2. Connect the other end to the on-site compressed air supply. A coupling connector (NW 7.2) is included in the scope
of delivery of the device.
Do not remove the sterile filter pre-fitted to the device.
WARNING
If the steam sterilizer reports the occurrence of event 42 “Drying – pressure pump” or a program is
aborted without the compressed air supply having been completed at this point, the program is
viewed as not having been completed successfully.
n In such a case, the Universal-Program S must be performed five times without a load (alternatively, three
times with activated additional drying). Drying must not be aborted. Only then is the steam sterilizer ready
for operation.
n The steam sterilizer can only be operated correctly with permanently connected compressed air supply.
n Before every program start, ensure that the compressed air supply shut-off valve is open.
Switching on the steam sterilizer
The following must be fulfilled or present:
ü
The feed water container is connected and filled with fresh feed water of appropriate quality. Alternatively, a water
treatment unit is connected.
ü
The wastewater container is connected.
ü
The compressed air line is connected and any on-site shut-off valve is open.
ü
The steam sterilizer power supply is switched on.
u Switch on the steam sterilizer at the power switch.
Ê The display switches between the initial state and the message Unlock door with '+' key, as long as the
door is closed.
ÊAfter switching the steam sterilizer on, it requires approx. 15 min to reach operational readiness, if the Autom. pre-
heating option is activated.
PLEASE NOTE
For hose lengths over 1.5 m, event 14 may occur several times during initial commissioning.
n Acknowledge this message until it is no longer displayed.
In case of insufficiently filled containers or an unconnected water treatment unit, a signal tone sounds and
a respective message is shown on the Display.