12 System malfunctions
55
Event Possible causes What you can do
Poor drying results The steam sterilizer is overloaded. In addition to an orderly device function, the
drying depends to a large extent on the correct
setup and loading of the steam sterilizer.
1. Comply with the maximum permissible load
quantities, see Selecting the program
[}page33].
2. Activate pre-heating, see Selecting auto-
matic preheating [}page33].
3. Use the additional drying function, see Se-
lecting additional drying [}page34].
Warnings
Event Possible causes What you can do
Attention: Door open and
Start not possible
The door contact is not closed during the
starting process.
Set the correct position of the twist grip – 90°
across the longitudinal axis of the steam
sterilizer. The display shows Door closed.
Attention: No feed
water / Feed water
Please refill – Start not
possible
When using the feed water container:
There is not enough feed water in the
container or the inlet hose/cable is not
connected correctly.
1. Check the water level in the feed water con-
tainer and top up with feed water if neces-
sary.
2. Check that the hose/cable connector is cor-
rectly connected to the feed water contain-
er.
The float switch is blocked. 1. Check if the float switch in the feed water
container is blocked or jammed.
2. Check the screen insert in the feed water
container for blockages.