3.3 Connecting a computer
■■■■■■■■■■■■■■■■■■■■■■
16
■■■■■■■■
855 Robotic Titrosampler
Connecting the power cord
Accessories Power cord with the following specifications:
■ Length: max. 2 m
■ Number of cores: 3, with protective conductor
■ Instrument plug: IEC 60320 type C13
■ Conductor cross-section 3x min. 0.75 mm
2
/ 18 AWG
■ Power plug:
– according to customer requirement (6.2122.XX0)
– min. 10 A
NOTICE
Do not use a not permitted power cord!
1
Plugging in the power cord
■ Plug the power cord into the instrument's power socket.
■ Connect the power cord to the power grid.
3.3 Connecting a computer
The 855 Robotic Titrosampler requires a USB connection to a computer in
order to be able to be controlled by a computer software. Using a
6.2151.000 controller cable, the instrument can be connected directly,
either to a USB socket on a computer, to a connected USB hub or to a
different Metrohm control instrument.
You need administrator rights for the installation of driver software and
control software on your computer.
Cable connection and driver installation
A driver installation is required in order to ensure that the 855 Robotic
Titrosampler is recognized by the computer software. To accomplish this,
you must comply with the procedures specified. The following steps are
necessary:
1
Install the software
■ Insert the computer software installation CD and carry out the
installation program directions.
■ Exit the program if you have started it after the installation.