EasyManuals Logo

Metrohm 855 User Manual

Metrohm 855
69 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #65 background imageLoading...
Page #65 background image
■■■■■■■■■■■■■■■■■■■■■■
7 Technical specifications
855 Robotic Titrosampler
■■■■■■■■
57
7.3 Membrane pump(s) with valve
Flow rate > 450 mL/min
Pressure head 2 m
7.4 786 Swing Head
Maximum load approx. 15 N
Swing rate 10 - 55 angular degrees/s
Beaker sensor
connector
Socket with M8 thread
7.5 Interfaces and connectors
Connector "Con-
troller"
USB upstream port (9-pin Mini DIN socket) for connecting a computer
for controlling the instrument.
MSB connectors
MSB1 - MSB3
Three 9-pin Mini DIN sockets for connecting dosing devices (Dosino/
Dosimat), stirrers, etc.
USB connectors
1/2
Two USB downstream ports (type A sockets), each 500 mA, for con-
necting Metrohm instruments or USB peripheral devices of other man-
ufacturers.
Stirrer connector DIN socket
Stirring rate 722/802 rod stirrer: 180 - 3000 rpm
741 magnetic stirrer: 180 - 2600 rpm
adjustable in 15 steps each in both directions of rotation
Pump connector Two sockets with M8 thread for 772 Pump Unit or 823 Membrane
Pump Unit
U = 16 ± 1 V, I = 0.8 A
Swing Head con-
nector
9-pin Mini DIN socket