7.2 Lift and turntable
■■■■■■■■■■■■■■■■■■■■■■
56
■■■■■■■■
855 Robotic Titrosampler
–1250 - –1210 mV / +1210 - +1250 mV: non-guaranteed values,
dependent on reference voltage +2.5 V
Measuring
range
–120 - +120 µA
Resolution 0.1 µA
7.1.3 Temperature
One measuring input (Temp.) for a Pt1000 or NTC temperature sensor.
Automatic temperature compensation for NTC sensors; R (25 °C) and B
value can be configured.
Measuring range
Pt1000 –150 - +250 °C
NTC –5 - +250 °C
(R (25 °C) = 30,000 Ω and B (25/50) = 4,100 K)
Resolution
Pt1000 0.1 °C
NTC 0.1 °C
Measuring accu-
racy
Pt1000 ±0.2 °C
(Applies for measuring range –20 - +150 °C; ±1 digit; without sensor
error, under reference conditions)
NTC ±0.6 °C
(Applies for measuring range +10 - +40 °C; ±1 digit; without sensor
error, under reference conditions)
7.2 Lift and turntable
Stroke path
235 mm
Maximum lift load approx. 30 N / 3 kg
Lift rate adjustable, 5 - 25 mm/s
Turntable speed adjustable, 3 - 20 angular degrees/s