EasyManua.ls Logo

Metrohm 855

Metrohm 855
69 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7.2 Lift and turntable
■■■■■■■■■■■■■■■■■■■■■■
56
■■■■■■■■
855 Robotic Titrosampler
–1250 - –1210 mV / +1210 - +1250 mV: non-guaranteed values,
dependent on reference voltage +2.5 V
Measuring
range
–120 - +120 µA
Resolution 0.1 µA
7.1.3 Temperature
One measuring input (Temp.) for a Pt1000 or NTC temperature sensor.
Automatic temperature compensation for NTC sensors; R (25 °C) and B
value can be configured.
Measuring range
Pt1000 –150 - +250 °C
NTC –5 - +250 °C
(R (25 °C) = 30,000 Ω and B (25/50) = 4,100 K)
Resolution
Pt1000 0.1 °C
NTC 0.1 °C
Measuring accu-
racy
Pt1000 ±0.2 °C
(Applies for measuring range –20 - +150 °C; ±1 digit; without sensor
error, under reference conditions)
NTC ±0.6 °C
(Applies for measuring range +10 - +40 °C; ±1 digit; without sensor
error, under reference conditions)
7.2 Lift and turntable
Stroke path
235 mm
Maximum lift load approx. 30 N / 3 kg
Lift rate adjustable, 5 - 25 mm/s
Turntable speed adjustable, 3 - 20 angular degrees/s

Table of Contents

Related product manuals