3.8 Connecting the eluent bottle
■■■■■■■■■■■■■■■■■■■■■■
26
■■■■■■■■
883 Basic IC plus
6
Mounting the adsorber tube
NOTICE
Depending on the eluent used, the adsorber tube (6.1609.000)
must be filled differently:
■ For alkaline eluents or eluents with a low buffer capacity: first a
little cotton, then with CO
2
adsorber material.
■ For all other eluents: only with cotton.
■ Remove the plastic cover from the large opening of the adsorber
tube. Fill the adsorber tube and close it again using the plastic
cover.
■ Insert the adsorber tube into the bottle cap's large opening. Fas-
ten it to the bottle cap using the SGJ clip (6.2023.020).