4.10 Suppressor
■■■■■■■■■■■■■■■■■■■■■■
80
■■■■■■■■
883 Basic IC plus
4.10.3.3 Cleaning the suppressor
In the following cases, it may be necessary to clean the suppressor:
■ Increased backpressure at the suppressor's connection tubing.
■ Irremediable blockage of the suppressor (solutions can no longer be
pumped through the suppressor).
■ Irremediable jamming of the suppressor (suppressor can no longer be
switched over).
Cleaning the suppressor
1
Disconnecting the suppressor from the IC system
■ Switch off the instrument.
■ Disconnect all capillaries of the suppressor from the IC system.
2
Dismantling the suppressor
■ Unscrew the union nut (25-1) from the housing (25-5).
■ Pull the connecting piece (25-2) out of the suppressor drive
together with the rotor (25-3).
If rotor A gets stuck in the suppressor drive, you can push it out as
follows:
Put a pointed object into the slot in the suppressor drive and use
it to push out the rotor A.
■ Loosen the connecting piece from the rotor.
■ Remove the rotor from the adapter.
3
Rinsing the capillaries
■ Connect each of the 6 PTFE capillaries connected to the connect-
ing piece (25-2) in sequential order to the high-pressure pump
and pump ultrapure water through them.
■ Check whether water comes out at the connecting piece.
If one of the capillaries remains blocked, the connecting piece (25-2)
must be replaced (order number 6.2835.010).
4
Cleaning the rotor
■ Clean the sealing surface of the rotor (25-3) with ethanol using a
lint-free cloth.