Problem Cause Remedy
The eluent path is blocked. Check the eluent path and eliminate the block-
age.
The eluent contains gas
bubbles.
Deaerate the high-pressure pump (see
chapter 4.18, page 50).
Baseline increasing
strongly.
Insufficient flow of regener-
ation solution or rinsing
solution.
See problem "MSM – Insufficient flow of
regeneration solution or rinsing solution."
MSM – Reduced capacity. Regenerate the MSM (see chapter 6.11.3.2,
page 84).
SPM – Reduced capacity. Regenerate the SPM (see chapter 6.11.3.2,
page 84).
Chromatograms
have poor resolution
Separation column –
Diminished separating effi-
ciency.
Regenerate the separation column (see
chapter 6.16.4, page 96).
Replace the separation column (see
"Connecting the separation column", page
56).
Extreme spread of
the peaks in the
chromatogram.
Splitting (dual
peaks)
Capillary connections –
Dead volume in the sys-
tem.
Check the capillary connections (see chapter
4.2, page 15) (use PEEK capillaries with an
inner diameter of 0.25 mm between the injec-
tion valve and the detector).
Guard column – Dimin-
ished efficiency.
Replace the guard column (see chapter 4.19,
page 52).
Separation column – Dead
volume at the column
head.
Install the separation column in the oppo-
site flow direction (if the leaflet says this is
permissible) and rinse into a beaker.
Replace the separation column (see
"Connecting the separation column", page
56).
Precision problems -
the measured values
are highly scattered.
Injection valve – Sample
loop.
Check the installation of the sample loop (see
"Optional: Exchanging the sample loop",
page 29).
Sample – There are gas
bubbles in the sample.
Use the sample degasser .
Sample – The rinsing vol-
ume is too small.
Increase the rinsing time (see chapter 6.15,
page 94).

Table of Contents

Other manuals for Metrohm 930 Compact IC Flex

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 930 Compact IC Flex and is the answer not in the manual?

Metrohm 930 Compact IC Flex Specifications

General IconGeneral
BrandMetrohm
Model930 Compact IC Flex
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals