6 Operation and maintenance
930 Compact IC Flex Oven/SeS/PP (2.930.2500)
■■■■■■■■
79
Connecting the entry and exit for the high-pressure pump
1
Remove the stopper from the coupling. Tighten the coupling to
the pressure screw located on the pump head inlet capillary (1).
Reconnect the pump head outlet capillary to the pump head out-
let (2).
6.8 Servicing the inline filter
NOTE
You can find a video sequence for this task in the Multimedia Guide IC
Maintenance or on the Internet at http://ic-help.metrohm.com/.
Maintenance interval
The filter must be replaced at least every 3 months; it may need to be
replaced more frequently, depending on the application.
Accessories For this task, you need the following accessories:
Two adjustable wrenches (6.2621.000) from the accessory kit: Vario/
Flex Basic (6.5000.000)
Tweezers
A new filter from the packaging (6.2821.130)