6 Operation and maintenance
930 Compact IC Flex Oven/SeS/PP (2.930.2500)
■■■■■■■■
65
The composition of the eluent plays a critical role in chromatographic
analysis:
Concentration An increase in the concentration generally leads
to shorter retention times and faster separation,
but also to a higher background conductivity sig-
nal.
pH pH changes lead to shifts in dissociation equili-
bria and thus to changes in retention times.
Organic solvents Adding organic solvents (e.g. methanol, acetone
or acetonitrile) to a watery eluent generally
speeds up lipophilic ions.
6.5.2 Changing the eluent
Ensure that no precipitates can form when changing the eluent. Immedi-
ately successive solutions must be miscible. If the system must be rinsed
with organic solvents, several solvents with rising or falling lipophilicity
must be used.
NOTE
To change the eluent, remove the guard column and the separation col-
umn. Connect the capillaries using a coupling (6.2744.040) and two
pressure screws (6.2744.070).
6.6 Notes on operating the high-pressure pump
CAUTION
The pump head is filled ex works with methanol/ultrapure water. Ensure
that the eluent used is miscible with this solvent.
Keep the following recommendations in mind in order to protect the high-
pressure pump from damage as much as possible during operation:
To protect the high-pressure pump from foreign particles, we recom-
mend filtering the eluent through a filter with a pore size of 0.45 µm
and aspirating it via an aspiration filter (6.2821.090).

Table of Contents

Other manuals for Metrohm 930 Compact IC Flex

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 930 Compact IC Flex and is the answer not in the manual?

Metrohm 930 Compact IC Flex Specifications

General IconGeneral
BrandMetrohm
Model930 Compact IC Flex
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals