6.8 Servicing the inline filter
■■■■■■■■■■■■■■■■■■■■■■
80
■■■■■■■■
930 Compact IC Flex Oven/SeS/PP (2.930.2500)
Removing the filter
Figure 23 Inline filter – Removing the filter
1
Filter housing
Inline filter housing. Part of the accessories
(6.2821.120).
2
Filter screw
Screw for the inline filter. Part of the acces-
sories (6.2821.120).
3
Filter (6.2821.130)
Pack contains 10 pieces.
1
Shutting off the flow
Switch off the high-pressure pump in the software.
2
Removing the inline filter
Unscrew both pressure screws from the inline filter.
3
Unscrewing the filter screw
Use two adjustable wrenches (6.2621.000) to loosen the filter screw
(23-2) from the filter housing (23-1) and unscrew it by hand.
4
Removing the filter
Remove the old filter (23-3) using tweezers.