8.1 Reference conditions
■■■■■■■■■■■■■■■■■■■■■■
102
■■■■■■■■
930 Compact IC Flex Oven/SeS/PP (2.930.2500)
8 Technical specifications
8.1 Reference conditions
The technical specifications listed in this chapter refer to the following ref-
erence conditions:
Ambient temp-
erature
+25 °C (± 3 °C)
Instrument status Operating > 40 minutes (in equilibrium)
8.2 Instrument
IC system
Metal-free IC system
Compact system with modular design
Material Painted, CFC-free polyurethane hard foam, fire class V0
Operating pres-
sure range
0–50 MPa (500 bar) high-pressure pump
0–35 MPa (350 bar) standard PEEK system
Intelligent com-
ponents
iPump, iDetector, iColumn, MagIC Net
8.3 Ambient conditions
Operation
Nominal function
range
+5–+45 °C
at max. 80% relative humidity, non-condensing
Storage +5–+45 °C