4 Installation
930 Compact IC Flex Oven/SeS/PP (2.930.2500)
■■■■■■■■
15
4 Installation
4.1 Setting up the instrument
4.1.1 Packaging
The instrument is supplied in protective packaging together with the sepa-
rately packed accessories. Keep this packaging, as only this ensures safe
transportation of the instrument.
4.1.2 Checks
Immediately after receipt, check whether the shipment has arrived com-
plete and without damage by comparing it with the delivery note.
4.1.3 Location
The instrument has been developed for operation indoors and may not be
used in explosive environments.
Place the instrument in a location of the laboratory which is suitable for
operation and free of vibrations and which provides protection against
corrosive atmosphere and contamination by chemicals.
The instrument should be protected against excessive temperature fluctua-
tions and direct sunlight.
4.2 Capillary connections in the IC system
Generally speaking, capillary connections between 2 components of an IC
system are made up of one connection capillary and 2 pressure screws
used to connect the capillary to the respective components.
Pressure screws
Three types of pressure screws are used in the IC system:
Number
Designation Use
6.2744.010 / 6.2744.014 Pressure screw On the injection valve
6.2744.070 Pressure screw, short High-pressure pump, purge
valve, inline filter, pulsation
absorber, separation columns
6.2744.090 Pressure screw, long MCS, sample degasser, 10-
port valve
Pressure screws are tightened and loosened by hand. A tool is not
needed.

Table of Contents

Other manuals for Metrohm 930 Compact IC Flex

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 930 Compact IC Flex and is the answer not in the manual?

Metrohm 930 Compact IC Flex Specifications

General IconGeneral
BrandMetrohm
Model930 Compact IC Flex
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals