6.12 Peristaltic pump
■■■■■■■■■■■■■■■■■■■■■■
92
■■■■■■■■
930 Compact IC Flex Oven/SeS/PP (2.930.2500)
Maintenance interval Replace the pump tubing every 2 months.
Replace the pump tubing every 4 weeks if the peristaltic pump is being
used continuously.
6.12.2.2 Replacing the filter
The filters inserted into the pump tubing connection with locking nut and
filter (6.2744.180) must be replaced regularly.
Maintenance interval We recommend replacing the filters (6.2821.130) (25-2) every 3 months.
The filters may need to be replaced more frequently, depending on the
application.
Accessories For this step, you need the following accessories:
■ 1 filter from the spare filter set (6.2821.130)
■ 2 adjustable wrenches (6.2621.000)
■ Tweezers
Replacing the filter
Figure 25
Pump tubing connection – Replacing the filter
1
Tubing olive
2
Filter (6.2821.130)
Pack contains 10 pieces.
3
Filter screw
1
Unscrewing the filter screw
■ Unscrew the filter screw (25-3) from the tubing olive (25-1) using
the two adjustable wrenches.
2
Replacing the filter
■ Remove the old filter (25-2) using tweezers.
■ Use tweezers to place the new filter (25-2) on the filter screw
(25-3) so that it is flat and press it firmly into place with the rear
of the tweezers.